Original Jpn J Radiol Technol 1999 ;
Analysis of Verification Image Using Electronic Portal Imaging Devices(EPID)

SHINICHI INOUE, MITSUHIRO MATSUMOTO, HIRONORI HASEGAWA, KAZUHIKO SATOH, KUNIAKI SAKURAI, YASUMITSU KAWAGOE, and YOSHIMI KUSUMI
Department of Radiology, Osaka University Hospital
Received Aug. 14, 1998; Revision accepted Feb. 22, 1999; Code Nos. 400, 410

Summary
We studied the evaluation of patient set-up error by verification image using electronic portal imaging devices(EPID). We analyzed which cleared 3 mm in our hospital for visual by using a distance measurement function. The body parts examined included the head and neck, thorax, pelvis and abdomen. The items that were evaluated included shell, Pb block, multi-leaf collimator and landmark. Patient set-up error was 10.4%, with that for the pelvis especially poor. The visual false rate was 17.9%. The result for the abdomen was poor because of the absence of landmark. We showed a decrease in patient set-up error by using EPID and improvement of radiotherapy .

Key words: Radiotherapy, Electronic portal imaging device(EPID), Verification